Next: Fringe Fitting
Up: PHASE CALIBRATION AND IMAGING
Previous: Fringe Finders
Contents
The Pulse Cal System
VLBA observers using more than 1 BBC will want to sum over the
BBCs to reduce noise levels. This should not be done with the raw
signals delivered by the BBCs: the independent local oscillators in
each BBC introduce an unknown phase offset from one BBC to the next,
so such a summation of the raw signals would be incoherent. A
so-called ``phase cal'' or ``pulse cal'' system (Thompson 1995) is
available at VLBA antennas to overcome this problem. This system, in
conjuction with the LO cable length measuring system, is also used to
measure changes in the delays through the cables and electronics which
must be removed for accurate geodetic and astrometric observations.
The pulse cal system consists of a pulse generator and a sine-wave
detector. The interval between the pulses can be either 0.2 or 1
microsecond. They are injected into the signal path at the receivers
and serve to define the delay reference point for astrometry. The
weak pulses appear in the spectrum as a ``comb'' of very narrow, weak
spectral lines at intervals of 1 MHz (or, optionally, 5 MHz). The
detector, located at the VLBA antennas, measures the phase of one or
more of these lines, and their relative offsets can be used to correct
the phases of data from different BBCs. The VLBA pulse cal data are
logged as a function of time and delivered in a PC table (see
Section 16). AIPS software can be used to load and
apply these data. However, some VLBA observers may still want to use
a strong compact source to do a ``manual'' pulse cal if necessary
(Diamond 1995). For example, spectral line users will not want the
pulse cal ``comb'' in their spectra, so they should ensure that their
observing schedules both disable the pulse cal generators and include
observations suitable for a ``manual'' pulse cal. Manual pulse
calibration also is likely to be necessary for any non-VLBA antennas
included in an observation, because they may have no tone generators,
or else may not have detectors located at the antenna. In addition,
it is necessary at 3 mm, where the VLBA antennas have no pulse
calibration tones.
Next: Fringe Fitting
Up: PHASE CALIBRATION AND IMAGING
Previous: Fringe Finders
Contents
Jim Ulvestad
2008-08-04