In recent months, some problems have been identified with the calibration transfer data distributed with VLBA correlator data. All of these problems affect only small amounts of data, and they do not affect the actual correlation results. Final processing results can, however, be adversely affected through the application of flagging and calibration. The issues are: 1. Occasionally the correlator job generation program creates spurious flags for the FG table with the reason "System idle". This is generally at the start of a job script and relates to the bookkeeping of what flags are set. It can cause major portions of a good scan to be flagged. That could be handled by deleting such flags, except while that flag is set, no data are being passed to the calibration tables so the data cannot be calibrated properly. The problem occurs sufficiently infrequently that not all projects are affected. The cure, if excessive data are affected, is to use the old data path for calibration using VLOG, UVFLG, ANTAB, and PCLOD after deleting the FG, GA, TY, and PC tables. The amount of affected data can be determined by looking at the FG table for the reason "System idle" or simply by running something like VPLOT with the flags on and off (FLAGVER=-1) and looking for suspicious data loss with them on. Remember that flags are on when FLAGVER=0. 2. It is possible for the calibration transfer system to mis-label data in the pulse cal (PC) table when a station that is in other correlator jobs for the project are not in one of the jobs. This can happen especially with MK and SC which cannot see sources for the full time that others are observing. What has been seen is that the data for the station that is missing is not simply thrown away, but rather is ascribed to another station. A plot of the pulse cal data with SNPLT or a listing of the data will show twice as many calibration points as should be there, and alternate rows will have different patterns of relative phases between IFs or polarizations. Inspection of the patterns will show that half the rows belong to the missing station. Running the pulse cal calibration (PCCOR) on such data would create badly calibrated data. Again, the workaround involves using VLOG and PCLOD. Or the offending points could be edited from the PC table. 3. The time associated with pulse cal data points needs some adjustments from the values in the monitor data base. Data in the cal file (VLOG input) has all known adjustments applied. But from calibration transfer, there can be a 10s offset for truncated integrations - typically integrations less than 2 minutes long. Since the pulse cal values generally change slowly within a scan, this will have very little affect on the data in practice and likely can be ignored. It is possible, but very unlikely, that it will also cause a point to appear to be after the scan in which it was measured, which could cause it not to be used in calibration. All of these issues potentially affect data from the start of the use of calibration transfer around late 1998, although the flag bug may be more prevalent since the use of Mark5 began. It is not clear when they will be fixed as the detailed causes are not yet fully understood and the people involved have higher priority work to do on the EVLA. --------------------------------------------------------------------- R. Craig Walker Array Operations Center cwalker@nrao.edu National Radio Astronomy Observatory Phone 505 835 7247 P. O. Box O Fax 505 835 7027 Socorro NM 87801 USA ---------------------------------------------------------------------